Description
This technique allows a broad band measurement of the complex permeability of isotropic thin layers. The samples under test are placed onto a dielectric holder.
|
|
Principle
This technique is based on the measurement of the impedance of a 2-plate line shunted at one end. The sample consists in a thin or thick, rectangle- or square-shaped layer placed under the line strip in front of the shunt.
|
|
|
Analysis of measured data
The shunted line constitutes an inductive loop. The complex permeability of the layer, µ*, is determined from the measurement of the inductive loop impedance. To discard the systematic errors inherent in the test devices, e.g. analyzer, line, connector.., a vector subtraction is made between the measurement and the one under load.
|
|
Résultats
- Frequency range used: 130 MHz - 8 GHz (network analyzers HP8720A).
- Type of material tested: isotropic magnetic thin or thick layers.
- Network analyzer calibration: SOLT.
- Dimensions of the tested samples:
- length: 9 mm,
- width: 9 mm,
- thickness: less than 1.8 mm including the substrate that holds the film.
|
Permeability spectrum of a 154-mm-thick CoNbZr thin layer onto a rigid substrate
|
|
Comments
The above figure displays the spectrum of a marked gyromagnetic resonance.
|
|